The idea of using light reflection instead of light transmission in clinical oximetry shows that SpO2 can be monitored by measuring the amount of light reflected from the tissue. The idea of using skin reflectance spectrophotometry marked a significant advancement in the noninvasive monitoring of SpO2 from virtually any point on the skin surface.
The intensity of the back scattered light from the skin depends not only on the optical absorption spectrum of the blood but also on the structure and pigmentation of the skin. SpO2 is measured be analyzing the pulsatile componens of the detected red and infrared plethysmograms which make use of reflected light intensities. The light from the LEDs enters the tissue, is scattered by both the moving red blood cells and the nonmoving tissue, and a part of this back scattered light is detected by the photodiode.
The output of photodiode is processed by the pulse oximeter, and measures the SpO2 of the pulsatile blood. In reflectance pulse oximetry, the LEDs and the photodiode are placed on the same side of the skin surface as shown below. Normally the reflectance probe is placed on the forehead or temple, but is not restricted to only those two places. Reflectance probes can be used to measure arterial oxygen saturation at virtually any place the human body where the probe can be placed.
DCM01 DCM02 developed by APMKorea is a sensor integrated built in LEDs and PD together with optimized properly against such trade-off to get proper parameters. Furthermore DCM01 DCM02 has SMD type package so that user could interconnect easily wire or flip bonding into your set.
This reflectance sensor can be placed on the body where you can expect light reflection due to tissue.
Features
- Integrated Dual Emitter and Photo Detector
- Dual Red and IR LED Matched to SPO2
- Chip on Board Package
Applications
- Reflective Oximeter Probes
- Blood Analysis
- Medical Instrumentation
- Radiometric Instrument
Absolute MAX. Ratings(Ta=25°C unless otherwise noted)
Symbol
|
Parameter
|
Max. Rated Value
|
Units
|
PD
|
Power Dissipation
|
120
|
mW
|
IF
|
Continuous Forward Current
|
50
|
mA
|
IFR
|
Peak Forward Current
|
100
|
mA
|
VR
|
Reverse Voltage
|
5
|
V
|
T stg
|
Storage Temp.
|
-20~+85
|
°C
|
T opr
|
Operating Temp.
|
-20~+85
|
°C
|
T sd
|
Soldering Temperature *1
|
240
|
°C
|
*1. Test Condition : t<3s |
Characteristics(Ta=25°C unless otherwise noted)
Symbol
|
Characteristic
|
Components
|
Test Condition
|
Min
|
Typ
|
Max
|
Units
|
VF
|
Forward Voltage
|
LED1
|
IF=20mA
|
1.9
|
2.2
|
V
|
|
LED2
|
1.3
|
1.5
|
|||||
PD
|
IF=10mA
|
0.2
|
1.3
|
||||
IR
|
Reverse Breakdown Current
|
LED1, LED2
|
VR=5V
|
10
|
uA
|
||
P O
|
Output Power
|
LED1
|
IF=20mA
|
2
|
mW
|
||
LED2
|
2
|
||||||
λ PEAK
|
Peak Wavelength
|
LED1
|
IF=20mA
|
650
|
660
|
670
|
nm
|
LED2
|
880
|
905
|
910
|
||||
PD
|
940
|
||||||
Δλ
|
Half Wave Width
|
LED1
|
IF=20mA
|
30
|
nm
|
||
LED2
|
60
|
||||||
V BR
|
Reverse Breakdown Voltage
|
PD
|
IR=100uA
|
35
|
V
|
||
I D
|
Reverse Dark Current
|
V R=10V
|
20
|
nA
|
|||
I L
|
Light Current
|
1mW, 940nm
|
45
|
uA
|
|||
S
|
Spectral Response Rrange
|
400
|
1050
|
nm
|
|||
CJ
|
Junction Capacitance
|
VR=3V, f=1MHz
|
44
|
pF
|
PPG(Photo Plethysmogram) Characteristic
B2B Trade
Price (FOB) | Negotiable | transportation | - |
---|---|---|---|
MOQ | Negotiable | Leadtime | Negotiable |
Payment Options | Negotiable | Shipping time | Negotiable |
- President
- shoon Han
- Address
- Noblesse201, Hanbat-daero138, Daejeon Korea
- Product Category
- Sensors
- No. of Total Employees
- 1-50
- Company introduction
-
President/owner shoon Han has founded APMKorea (Advanced Prime Materials) 9th, September 2008.
He has experience for 25years in semiconductor industry.Joined the member of the KITA (Korea International Trade Association) Register No. 16118892.
Shipped SPO2 blood sensor chips for medical device customers in Germany, India, Indonesia, Singapore, Turkey and Korea. June 2009~.Developed reflective blood sensor chips, providing new idea and varial applications for medical device customers in Singapore, Germany, Ukraine, India. March 2010~.
Joined New Exporters 300 as a member, consulted by KITA (Korea International Trade Association). 23rd March 2010.
Shipped SPO2 blood sensor chips for medical device customers in Switzerland and Brazil May 2010~.Registered a patent for the reflective blood sensor, 10-2010-0047289 20th May 2010.
Shipped SPO2 blood sensor chips for medical device customers in Colombia 20th May 2010~.
Presented at the Technology Road Show in Shanghai & Beijing China 6~10th Sep. 2010
Shipped Refective SPO2 blood sensor to Japan 15th Sep. 2010~.Shipped SPO2 blood sensor chips 8000pcs for medical device customers in Shanghai China Dec 3rd 2010~.
Shipped SPO2 blood sensor chips for medical device customers in Uruguay 11th Dec 2010~.
Shipped SPO2 blood sensor chips for medical device to India, Israel, Vietnam, US, New zealand, UK, France, Slovenia, Bulgaria, Italy 2011~.
Transmissive SpO2 Oximeter sensor qualified and supply to Germany June 2011~
Shipped SpO2 oximeter sensor to NASA (Johnson Space Center) on 30 June 2011.
- Main Markets
-
Brazil
Germany
Singapore
Switzerland
Ukraine
- Main Product
Related Products
Cable and Wire harness tester (MHT-610 / 705)
26/35kV -XLPE Insulated Wire Armoured Cable (XLPE / SWA / PVC)
Supvan TP76E tube printer, cable id printer, wire marking
18/30(36)kV-XLPE Insulated Unarmoured Cable (XLPE / PVC)
AlphaO DT-42 Non-stripping Wire Connector