Product Information
NUCLOCK Hot Cell series
The NUCLOCK hot cell series are all designed to accommodate dispensing systems or synthesis modules in safe, reliable, and sterile conditions. The NUCLOCK series with single or double either horizontally or vertically built is to have different configurations in regard to overall dimension, shielding requirement and various accessories.
The supportive material used for the structure of all the NUCLOCK is steel plate. The internal working area is made of stainless steel with well-polished surface, designed to allow for easy cleaning and decontamination.
The NUCLOCK series are all equipped with a touch screen that provides clear images, easy-to-use, and ergonomics, leading to accessing all the functions to be managed and controlled easily at a single point.
The NUCLOCK is also equipped with air tight system with acrylic glass panel held by inflatable gaskets mounted around the perimeter of each cells.
Vial extraction system for the NUCLOCK series is placed in a shielded container on the side of a cell where it can house a vial. It is therefore possible to extract the radiopharmaceuticals without opening the main door.
The ventilated pre-chambers for the introduction of materials are mounted on either side suitable for preparing dose of radio medicines.
Rsearch Hot-cell
The NUCLOCK N3 series aimed for research experimentation purpose are designed to house synthesis modules with the highest flexibility suitable for the new synthesis processes set up
The N3 cells are equipped with manipulation, plexiglass panel with gloves, mobile shield with lead glass window. All above mentioned features allow to use the cell in different configurations and in particular the manipulation with front lead door opened but in air tight conditions (with plexiglass panel closed)
Ventilation System
→ Inlet HEPA filters
→ Outlet Charcoal filters
Specification
Shielding |
Lead thickness : 50, 75, 100mm on all sides |
Internal view |
CCTV or Lead glass window 350(W) x 350(L) on main door |
Ventilation |
- Air quality : Class A, B, C - Inlet filter : HEPA (EI≥99.97%) - Outlet filter : Charcoal |
Dimension & Characteristics |
- External dimensions : 1400(W) x 1300(D) x 2700(H)mm - Inner dimensions : 850(W) x 850(D) x 800(H)mm - Weight : about 9000kg - Inlet Air : 4kg/cm2 - Inlet N2 : 7kg/cm2 - Vacuum : 400mmHg |
Utility |
Three phase 380V, 5kW / Single phase 220V, 2kW |
Main Door |
Up-down type Automatic up-down mobile door activated by motor installed low part of cells Swing type Hinged shielding door |
Option |
- Mobile shield with lead glass window - Dose - lift system - Tong with booting - Hand Glove - Pre-chamber |
* All the specifications illustrated above shall be tailored so as to fit into requests made by end-users. |
B2B Trade
Price (FOB) | Negotiable | transportation | - |
---|---|---|---|
MOQ | Negotiable | Leadtime | Negotiable |
Payment Options | Negotiable | Shipping time | Negotiable |
- President
- Choung Youngsub
- Address
- 40-19, Ori-ro 22da-gil, Guro-gu, Seoul, South Korea 152-883
- Product Category
- Beauty & Personal Care,Medical Devices
- Year Established
- 1991
- No. of Total Employees
- 1-50
- Company introduction
-
Every member of the company is dedicated to the development of nuclear medicine equipment in the pursuit of protecting and saving lives from being contaminated by hazardous materials. We Nuclear Light Ind. Co. Ltd., hereafter called NL, do believe that personal safety is the most valuable asset that we are pursuing. Our mission is therefore to give the people comfort, relief, and happiness by producing and supplying reliable devices that are designed and developed with a sophisticated manner.NL was founded in 1992 with the aim to enhance development of a variety of businesses related to the nuclear medicine, starting off with radiation shielding facility construction, documentation service for the approval of facilities where they produce radiopharmaceuticals, and supplying dose calibrators, and later extended into the development of the hot cell with its subsidiary devises, leading to one of prominent organization recognized by the market as an innovative firm in S. Korea.With nuclear medicine related equipment and well established know-how - piled up over two decades - of the development of radiation shielding technologies, NL has contributed to the industry in terms of radiation safety concerns and comfort use, resulting in the industrial enhancement with the accumulated radiation technique and experience. As the industrial leader, NL is committed to providing better service by continuously putting a step further toward clients.
- Main Markets
-
Jordan
Thailand
- Main Product